ZK 390-2 Endo-Urology 09.2019
Specification IM600-600 Irrigation tube sets with RFID (radio-frequency identification) ED 48 Endo Device RUDOLF Medical ZK 583-02 EN_Rev. 02.2019 Dimensions Dimensions (w x h x d) 380 x 148 x 388 mm Weigth 6.3 kg Electronics Protection class Protection class1 Application part acc. to IEC 60601-1 Type BF Protection type IP 41 Power supply 100V (AC) – 240V (AC), 50/60 Hz Power consumption 80 VA Current consumption 100V (AC), 0.8 A (80 VA), 240V (AC), 0.3 A (72 VA) Pump Data Indications covered Laparoscopy, Arthroscopy, Hysteroscopy, Urology Pressure range Laparoscopy: not adjustable Arthroscopy: 5 – 150 mmHg Hysteroscopy: 15 – 150 mmHg Urology: 10 – 90 mmHg Overpressure prevention Stringent safety measures reliably prevent overpressure in the application area Flow range Laparoscopy: 1.0 – 1.8 l/min Arthroscopy: 0.1 – 1.8 l/min Hysteroscopy: 50 – 500 ml/min Urology: 25 – 500 ml/min Suction capacity max. 2.0 l/min (depending on instrument characteristics) Display Easy to use Color touchscreen Connectivity The pump reacts to RFID protected tube sets and ensures proper use MULTI INDICATION PUMP ( STANDARD = HYSTEROSCOPY)