ZK 460-2 Laparoscopy 07.19
Dimensions Dimensions (w x h x d) 380 x 142 x 370 mm Weight 10 kg Electronics Protection class Protection class1 Current consumption 1.0 – 0.42 A Application part of the type (B, BF, CF) Type BF Protection type IP 21 Power supply 100V (AC) – 240V (AC), 50/60 Hz Fuse designation T3.15AH 250 V Power consumption 150 VA Current consumption 100V (AC), 60 Hz: 0.45 A (45VA), 240V (AC), 50 Hz: 0.25 A (60VA) Insufflator Data Insufflation medium Medical grade CO 2 Maximum output pressure 75 mmHg Maximum gas supply pressure 80 bar / 1160 PSI Min. gas supply pressure (bottle) 15 bar / 218 PSI Min. gas supply pressure (central gas) 3.4 bar / 50 PSI Max. gas flow Bariatric 50 l/min Standard 40 l/min Pediatric 20 l/min Vessel Harvesting 10 l/min Pressure range 1 – 20 mmHg (Pediatric/Vessel Harvesting) 1 – 30 mmHg (Standard / Bariatric) Modes Three additional modes Bariatric, Pediatric and Vessel Harvesting Gas heating Reusable and disposable Connectivity Tube sets Standard insufflation tube sets Gas supply “US” type connector for use with gas bottles exchangeable into: DISS or NIST for use withcentral gas supplies Specification IM600-500 ED 34 Endo Device RUDOLF Medical ZK 583-02 EN_Rev. 02.2019 CO 2 INSUFFLATOR 50 L /MIN