ED 9 Endo Device RUDOLF Medical ZK 583-02 EN Rev.02.2024 Specifications IM500-401 CAMERA HEAD 4K UHD ZOOM Technical Data Dimensions (w x h x d) 39.4 x 40 x 134 mm Weight of camera head (without cable) 330 g Resolution Camera output 3840 x 2160, 50/60p Angle of camera cable Straight camera head with approx. 40° angled camera cable Zoom Integrated 2x optical zoom Parfocal Optical zoom is parfocal Pixel error correction Dynamic pixel error combination. Pixel errors are automatically corrected by the camera system Sensor technology 3 x 1/3" CMOS, progressive scan 3-Chip sensor Digital data transmission Digital image signal through camera head cable Sterilizable Sterilizable with STERRAD 100NX Camera control functions 3 direct access buttons for sterile field control of the following functions: a. White balance b. Brightness c. Shutter d. Light source on/off e. Light source intensity up / down f. Digital special imaging algorithms g. Digital smoke reduction h. Grid removal i. Edge enhancement Buttons Illuminated camera head buttons Communication interface Communication interface to control light source settings via camera head CF rating The camera head is rated CF defibrillation proof according to EN 60601 allowing cardiac surgery Defibrillation proof rating The camera head is defibrillation-proof according to EN 6060 Cable length Approx. 3.5 m Endocoupler Standard eyepiece adapter Permanently connected (no C-Mount) Endoscope clamping lockable
RkJQdWJsaXNoZXIy ODU5MTg=